Notes de Rachel/Interns, Summer 2018 (51 notes)/Week 6
Jump to navigation
Jump to search
Notes de Rachel / Interns, Summer 2018 (51 notes)
27.08.18
- Continuation of the prototype:
- Laser cutting of a new foil made of acrylic
- =/= the machine is broken
- Laser cutting of a new foil made of acrylic
- Editing of the wiki
28.08.18
- Laser cutter works (a bit): we made a plate with very small holes
29.08.18
- MicroChip: Test the plate with very small holes for making the pads
- (The one to pour the gel into and from which it will go out and which will enable smooth surfaces for the pads)
- Works well, the surfaces are really smooth, except for the holes from which we put the gel, which are slightly visible
- Blue LED to test (given by Masoud) on a plate having SYBR safe
- Nothing to see, maybe the SYBR Safe was bleached
- Indeed, with the previous LEDs, we didn't see anything
- With new staining with SYBR Safe, we didn't see anything again
- Phosphorescent beads given by Masoud:
- Excitable by UVA (400 nm)
- Laser engraving from the FabLab (with Richard)
- Resolution of 10 micrometer?
- Not bad:
30.08.18
- Making of new solutions
- 0.5M Tris OK
- 0.5M EDTA pH 8 DONE
- We have 0.6679g (4.2037 g for a Falcon 15 ml) of 372.24 g/mol
- Therefore we can make 3.6 ml of EDTA (as 186.1g gives 1L)
- 5M NaCl or 2.5??? OK
- FINAL CONCENTRATION OF 2.5, so everything is fine
- 20180802 First Comet Assays, it's 2.5M (
- PK Buffer and PK+ Buffer, it's 5M (8ml)
- 10% Triton
- Comet assay
- By testing distilled water (from the Migros)
- Making of 0.5M EDTA pH 8
- Put 0.6679g of EDTA in a 15 mL Falcon
- Put 2 mL of Sterile Water
- (*)After calculations, we see that we need "en tout" 360 ul of NaOH
- For 1mL of EDTA, we need 20mg of NaOH pellets
- NaOH 5M = 200g/L = 200 mg/ml
- Which is equivalent to 20 mg in 0.1 ml
- So 0.1 ml of NaOH for 1 ml EDTA
- Therefore, we need 0.36 ml of NaOH for 3.6 ml of EDTA
- Put 5 ul of 5M NaOH OR 50 ul (I think I put 50 ul and 2.5 ul after)
- Put to shake on a "hot plate" (the one used to heat small tubes), for 1 hour at 70°C
- After 30 min, took 2 ul of the solution to test the pH: with the eye, we see a pH of ~5
- Put 2,5 ul of 5M NaOH
- After 30 min, took 5 ul of the solution to test the pH: with the eye, we see a pH of ~5
- Put 20 ul of 5M NaOH
- Calculations made see(*)
- Put 100 ul of 5M NaOH
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~6
- Put 75 ul of 5M NaOH
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~7
- Put 75 ul of 5M NaOH
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~7.5
- Put 25 ul of 5M NaOH
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~7.5
- Put 25 ul of 5M NaOH
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~7.5
- Here we have 327.5 ul OR 372.5 ul of NaOH (I think 372.5 ul)
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~8
- We have 2335.5 ul
- Put 1264.5 ul of sterile water, to have a final volume of 3.6 ml
- Took 5 ul of the solution to test the pH: with the eye, we see a pH of ~8
31.08.18
- Testing of different filters and lights with the spectrophotometer
- New test with SYBR Safe:
- We did a new spread of Vio's cells
- 100 ul of the cells
- "New" SYBR Safe (it's not the first time that we use it), in the black tube, at room temperature
- 4 ul of SYBR Safe was put directly on one part of the smear (without methanol and without washing the salts)
- We see fluorescence on two cells!
- The Big Filter: Let SYBR emission pass through, but the difference with the yellow filter alone is not huge
- --> The yellow filter is absolutely crucial, but we don't know why, as the spectre of light that it let pass does not help to distinguish the cells
- 4 ul of SYBR Safe was put directly on one part of the smear (without methanol and without washing the salts)