Notes de Rachel/Covid-19 LAMP tests (46 notes)/real time Isothermal Amplification

From Hackuarium
Jump to navigation Jump to search

Notes de Rachel / Covid-19 LAMP tests (46 notes)

To make a basic protocol for others to use:

For making a master mix for  x12 reactions:
for example, might have samples as below to test 
1,2 - NTC (non-template control, i.e. water),
3,4 - Positive control
5,6 - unknown 1
7,8 - unknown 2
9,10 - unknown 3
11,12 - unknown 4

Note: should determine best primer ratios and Mg++ levels empirically!
(i.e. 1.6uM Fip/Bip, 0.2 F3/B3  0.4 LF/LB  and 7mM Mg++ works very well for some primers)
Primer mix is made from 100mM stocks 20ul each FIP/BIP, 2.5ul each F3/B3 and 5ul LF/BF used in master mix as below

Per 20ul reaction:                (x13 for all components for the 12 reactions)         
2.0ul 10x buffer (i.e. isothermal buffer)                                                     (26ul)
1.1 ul primer mix                                                                                                 (14.3ul)                       
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction)   (13ul)      
2.8ul dNTP, 10mM                                                                                               (36.4ul)
0.8ul Polymerase (i.e. Bst2.0 (8units))                                                          (10.4)
1.6 ul SYBR-safe diluted 1/250 from stock solution.                              (20.8) 
water + sample to 20ul final   (say 2ul of N-dil or water per reaction)
         WATER  = 8.7 x13 = 113.1

After Master mix is made:

  1. distribute to the 12 tubes
  2. add ntc and close lids
  3. add samples closing lids as you go
  4. add positive control last (best practice is to not even take this tube out until master mixes are made and distributed...)
  5. run reactions in real time machine with FAM channel detection

For the old real time machine at Hackuarium, to have approximately Isothermal conditions, and also get good traces, this is the program used:
65 2min, 64 1min, then 'cycle' 10s 65 and 1:50s 64. (for imaging every 2 min)
Set for 35 cycles (with dissociation protocol clicked and 60 degrees...)
So, total times in minutes = (3 + Ct*2)

For a sample with RNA as input, of course need reverse transcription.  We use a 'one pot, two step' reaction, with the enzyme RTx for 10min at 55 before going onto the 64 cycling

Results: