To find out if the calibration of the big beast (based on Fluorescein and sybr-safe solution for background) will work

Try 20ul reactions
x2 water and x2RNA each (hope there are still 8ul of the original mRNA, if not just do Bst2.0 rxns)

Master Mixes
2.0ul 10x buffer   ThermoPol for oldschool Bst    Isothermal for Bst2.0            (x4.5 for all components)                                                                                                                                                    (9ul)
1.12ul* standard primer mix                                                                                 (5ul)
0.8ul extra MgSO4, 100mM (for 4mM more - 6mM final in reaction) (3.6 for all tubes)
2.8ul dNTP, 10mM                                                                                                  (12.6ul)
0.8ul Bst2.0 (8units) or BstOldSchool                                                                  (3.6)

1.6 ul SYBR-safe diluted 1/250 from stock solution, so final concentration is slightly decreased (0.00032), vs last time (1/2500=0.0004).                                                                                         (7.2 for all master mix tubes) 
water + sample to 20ul final   (say 2ul of RNA or water per reaction)

                                            20-2-1.2-.8-2.8-.8-1.6-2= 8.8. X4.5= 39.6 per master mix

Isothermal conditions trick ? up a few degrees each way?? hope!

In the end, annulling my test run on the big beast that looked so good, complicated everything, and the tubes got warmed then sat on the bench as I reset everything...
and, at least 4 of the tubes did have increased signals over the rest...
however, when playing with threshold as recommended by Guy, I hit the analyse button too soon, and could not seemingly go back!  hoping for help from Olivier!

May still need to redo calibrations!  Fluorescein alone is not how it is meant to be.
The end point picture had no obvious differences between signals, not like the last time with old school Bst, at least.