From RA's past experiences (most using Optigene reagents**):
A 'standard mix' of LAMP primers (as in Resuspension page 24.4.20) was used with
1.4ul in 25ul reaction to have these final concentrations:
1.6 uM FIP/BIP     
0.8 uM LB/LF        
0.4 uM F3/B3       

for very good primer sets, 0.7ul in 25ul reactions was used for half the above concentrations.

How does this compare to current LAMP scenarios and how much sybr safe should I try as intercalating fluor indicator?

The basic NEB protocol is found here:
https://www.neb.com/protocols/2014/12/29/typical-lamp-protocol-m0538

This recommends (FINAL CONCENTRATION !)

1.6uM FIP/BIP
0.4uM LB/LF
0.2uM F3/B3
(Their 25x mix is 40uM FIP/BIP, 10uM LB/LF and 5uM F3/B3)

-for first tests, need to decide best conditions!

These primer conditions are generally followed for the SYBR Safe paper which uses NEB enzyme etc.
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5568651/

(by Carter et al, and includes an Aronoff-Spencer)
From their methods:

Reactions were performed in 25 µL volume unless otherwise indicated in the text, and were comprised of 20 mM Tris-HCl, 10 mM (NH4)2SO4, 10 mM KCl, 8 mM MgSO4, 0.1% Triton X-100, 0.8 M betaine, 1.4 mM dNTPs, 1.6 µM each FIP/BIP primers (PAGE purified), 0.2 µM each F3/B3 primers, 0.4 µM each LoopF/LoopB primers, 0.32U/µL BST2.0 (New England Biosciences (NEB), Ipswich, MA, USA), 0.6U/µL RTx (NEB). Reactions were incubated at 63–65 °C for the indicated time, 60 min for most experiments.

Readout (fluor readings every 2 minutes) with Sybr-Safe at a final concentration of 10 × (a 1:2500 dilution of the manufacturer’s stock solution).

  • Here is what I had in a previous note from 'back when'

Our current primer optimisation does not test explicitly differing primer ratios, particularly of the outer to inner primer pairs,

In my experience, we mainly changed overall concentrations 0.5x, 1x, 2x in the reaction.
(with '1x' as 2.5ul each F3/B3, 5ul each LB/LF, and 10ul FIP/BIP in the standard primer mix)
as in an old SOP: This comes out to 1.6uM FIP and BIP,  0.8uM LB and LF, and 0.4uM F3 and B3
In comparison, Notomi et al (2000) used 0.8uM each FIP and BIP & 0.2uM each F3 and B3 (that was before the loop primers were found to help so much); and normal PCR usually has primer concentrations of 0.1-0.5uM.
doubling the loop primer mix is one way to try to optimise primer sets, but...

The correct way to determine the best ratios, would just starting with the F3/B3 to FIP/BIP pairs at different ratios:
(1:2, 1:4, 1:6; 1:8; 1:10)
5 different reaction mixes can be made, and tested in duplicate with the same amount of extract.  

For example 
see Wang et al (2016) . 
DOI 10.1007/s00216-016-0084-x .

Figure 3B shows their results, with 1:8 chosen for their primers...
(in comparison to using 1:4 in 'standard' mixes)

(To note: optimisation for Mg++ and nucleotide concentrations etc is also of use for optimisation, but in my previous experience OptiGene chose in that regard...) 

    • Here is some recent Optigene info

www.optigene.co.uk/gspm3-0-lf-dna-polymerase/

and here they use their primers at 0.8Fip/Bip, 0.4LB/LF and 0.2F3/B3

Choices need to be made.