Notes de Rachel/Covid-19 LAMP tests (46 notes)/280420 First RT-LAMP attempts

From Hackuarium
Jump to navigation Jump to search

Notes de Rachel / Covid-19 LAMP tests (46 notes)

Make Cellular RNA with the Monarch test kit
RNA1 with 10* 6 cells 
RNA2 with 10* 7 cells
(in the end, my counts of cheek cells showed only 5x105 per ml, with 7.5ml total, and I pelleted, so RNA2 had more cells - but spilled some sample before DNase steps... 
after reactions, should still have about 40ul left of each RNA sample)

Set up master mixes for 6 reactions each

  1. rActin primers
  2. RNaseP primers
  3. NM primers
  4. NZ primers

duplicate tubes for all 
0,0, RNA1 x2, RNA2 x2

For Bst2.0 +RTx reactions just try:
(for each 25ul reaction)
2.5ul 10x buffer                         (15)
1.5ul appropriate 'standard' primer mix (9)
1ul extra MgSO4, 100mM (for 4mM more - 6mM final in reaction) (6)
3.5ul dNTP, 10mM  (21)
1ul Bst2.0 (8units)  (6)
1ul RTx (15units) - to confirm on tube!  -- yes all good (6)
2.5 ul SYBR-safe diluted 1/250 from stock solution, so final concentration is 1/2500. (15) 
water + sample to 25ul final   (say 2.5ul or RNA or water per reaction? then 9.5ul...  (57)

(x6 for all)

to take pictures with CCD of ABI Prism7000 machine at least, if no real run possible
(every 3-5minutes?)
can also use UV to look at reactions (tube holder as in GMO detective)

Results:
    rActin up first, about 25min to product, but one water came up in these conditions.
    RNaseP up nicely also, about 32min - and waters stayed clean.
  NM and NZ stayed clean up to 45min, but one NM sample came up by 48min (and more  were showing signal by end)

     nice and bright positives vs negatives on trans-illum with blue led and amber filter
         (but could complain, it is not altogether 'black and white' esp NM tubes...)
    ready for tests with some positive controls