To find out if 4mM Mg++ and alt primers are even better for the NM primer set
Comparing to NZ - use alt primer mix

A as yesterday with 6mM Mg++
B with 4mM Mg++

x2 water (tubes 1,2/7,8) and x2 N plasmid dilutions ('100' - tubes 3,4/9,10 and '500' - tubes 5,6/11,12)
- for N control DNA possibly really 10 and 50 copies... (use 2ul of '50' for '100' and 1ul of '500' from original dilutions)
so 6 tubes for each condition/primer set
A - as yesterday's C for primer mix NM (1-6) and NZ (7-12), 6mM Mg++
B - with 4mM Mg++
So, x4 master mixes to make A1, B1 for NM, and A2, B2, for NZ

using only NZ and NM 'alt2loop' primer sets 

1.6uM FIP/BIP 
0.4uM LB/LF - actually 0.8! 
0.2uM F3/B3 

Master Mixes
2.0ul 10x buffer -  Isothermal buffer for Bst2.0           (x6.5 for all components for the 6 reactions)                                                                                                          (13ul)
1.12ul* alt primer mix                                                                                                                                                                                                                                          (7.3ul)                                                                     
0.8ul extra MgSO4, 100mM (for 4mM more - 6mM final in reaction)                                                                         OR                                                                                 (5.2ul for A master mix)
0.4ul extra MgSO4, 100mM (for 2mM more - 4mM final in reaction)     
                                                                                       (2.6ul for B)   
2.8ul dNTP, 10mM                                                                                                                                                                                                                                              (18.2ul)
0.8ul Bst2.0 (8units) or                                                                                                                                                                                                                                       (5.2)
1.6 ul SYBR-safe diluted 1/250 from stock solution.                                                                                                                                                                                         (10.4) 
water + sample to 20ul final   (2ul of N-dil or water per reaction with 18ul of appropriate master mix) 
                                            20-(2-1.12-.8-2.8-.8-1.6-2)= 8.88 X6.5 = 57.7 for A master mix
                                                                                                    60.3ul water for B master mix  
Again- Isothermal conditions trick for ABI to run and not complain up a few degrees each way, as worked yesterday:
65 2min, 64 1min, then 'cycle' 10s 65 and 1:50s 64. (for imaging every 2 min)
do 35 cycles again with dissociation protocol clicked and 60 degrees...

Results:

Only 3 tubes really amplified!  all NM (made pdf in thinkpad, but figure out easier data export etc, as most people won't be interested in the .sds files!)

4mM most likely not enough Mg++, but the 6mM should have been ok...
-> maybe the N plasmid dilution is not stable in water...
(to check if it should be diluted fresh for each experiment)

To re-do at least part of this