Notes de Rachel/Covid-19 LAMP tests (46 notes)/13102020 N-transformant test
Notes de Rachel / Covid-19 LAMP tests (46 notes)
To find out if the DH5alpha transformants got the N plasmid aok...
two colonies frozen in 20ul water 9oct.
(N1 lysate more dilute than N2 lysate)
master mixes x12 (1,2 - water, 3,4 N1lysate, 5,6 N2lysate; 7,8 N2 1/10dil 9,10 N2 1/100dil, 11,12 old N plasmid '1000'
Master Mixes
2.0ul 10x buffer - Isothermal buffer for Bst2.0 (x13 for all components for the 8 reactions) (26ul)
1.12ul* double loop alt primer mix as last time (14.6ul)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) (13ul)
2.8ul dNTP, 10mM (36.4ul)
0.8ul Bst2.0 (8units) (10.4)
1.6 ul SYBR-safe diluted 1/250 from stock solution. (20.8)
water + sample to 20ul final (say 2ul of N-dil or water per reaction)
WATER 18-9.32 = 8.7 x13 = 113.1
Again- Approximately Isothermal conditions, so get good traces:
65 2min, 64 1min, then 'cycle' 10s 65 and 1:50s 64. (for imaging every 2 min)
do 35 cycles again with dissociation protocol clicked and 60 degrees...
Results: