Notes de Rachel/Covid-19 LAMP tests (46 notes)/07082020 Multiplexes NMFAM-RPHex
Notes de Rachel / Covid-19 LAMP tests (46 notes)
To find out if the 10min extra for RTx was necessary for the good result with Bst2.0
(also repeat with BO again)
First clean with 10% bleach. Then check that I have enough primers for all sets of tubes (make more if needed) and RNAs
Master mixes for 8 rxn tubes (1,2 - water, 3,4 BEI RNA 100, 5,6 cell RNA, 7,8 BEI RNA 100 + cell RNA)
A: BO +RTx and RNAsin
B: Bst2.0 + RTx and RNAsin
x2 (so 4 8-tube strips: A+ and B+ with extra 55degree 10min)
Master Mixes for tagged primers:
2.0ul 10x buffer - Isothermal buffer for Bst2.0 (x17 for all components for the 16 reactions) (34ul)
1.12ul* tagged primer mix (19ul + 19ul) (DOUBLED for multis)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) (17ul)
2.8ul dNTP, 10mM (47.6ul)
0.8ul Bst2.0 (8units) or BstLarge Fragment (13.6)
for both master mixes RTx (13.6)
RNAsin (40units/ul) (0.5ul/reaction) (8.5)
water + sample to 20ul final (say 2ul of RNA or water per reaction)
18-(all components above -doubleprimers) 7.86x17= 134
NM fluor primer mix _FAM-NM*FIP + RNaseP fluor primer mix. _hex-RP*LF
Both quenchers 8.8ul for NM-FIP quench +4.4 for RP-LF quench, for water -> 120.5ul
- For pcr machine - start with 10min at 55 degrees for *+* tubes, to give RTx a better chance to work (since low!)****
- then put 0 tubes in too.****
- 1st picture after that + 5min of 65oC****
- then put 0 tubes in too.****
- always do A, A+, B, B+ for order of pics of tubes****
- by 35 min all the B+ + look good!****
- saw diff of 100copies - came up after the cell RNA signal today.****
- by 35 min all the B+ + look good!****
- is the diff more than 10min, without the RTx at 55degree step??****