Notes de Rachel/Covid-19 LAMP tests (46 notes)/05062020 NM8 with BEI RNA
Notes de Rachel / Covid-19 LAMP tests (46 notes)
To see how low we can detect the RNA...
dilution further and one test with saliva too...
2 master mixes x12 (1,2 - water, 3,4 BEI RNA 20, 5,6 BEI RNA 50; 7,8 BEI RNA 100, 9,10 BEI RNA 200, 11 BEI RNA 200 + Saliva and 12 N plasmid '1000'
A BO +RTx with NM
B Bst2.0+RTx with NM
Master Mixes
2.0ul 10x buffer - Isothermal buffer for Bst2.0 (x13 for all components for the 8 reactions) (26ul)
1.12ul* double loop alt primer mix as last time (14.6ul)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) (13ul)
2.8ul dNTP, 10mM (36.4ul)
0.8ul Bst2.0 (8units) or BstLarge Fragment (10.4)
for both master mixes RTx (10.4)
1.6 ul SYBR-safe diluted 1/250 from stock solution. (20.8)
water + sample to 20ul final (say 2ul of N-dil or water per reaction)
WATER with RTx-0.8 = 7.9 x13 = 102.7
Again- Approximately Isothermal conditions, so get good traces:
65 2min, 64 1min, then 'cycle' 10s 65 and 1:50s 64. (for imaging every 2 min)
do 35 cycles again with dissociation protocol clicked and 60 degrees...
Results:
messed up BO tubes so all got saliva in the end!
but some ok results...