Notes de Rachel/First Notebook (126 notes)/20240110 Nanopore run with Yngvar

From Hackuarium
Revision as of 20:55, 18 November 2026 by DavidMontavon (talk | contribs) (1 revision imported: Import des notes de Rachel — lot 3/6)
Jump to navigation Jump to search

Notes de Rachel / First Notebook (126 notes)

To make sure things are in working order, following the Kit14 sequencing strategy...

will need to do end repair and ligation of the q'gen DNAs

after each enyzme step (end repair d, repurify DNA ...

Yngvar thinks trying just one DNA is best to start, and do have AMPure beads in the kit!
so can do the purification with them, instead of qiagen!

tailed B1/BP1 (from previous experiment)  is the longer one Lactarius luridus
and will have the internal control DCS from the kit

have nuclease free water from Promega
Thaw the DCS and BP1 and put in ice
Ultra II End prep buffer and Enzyme on ice

Reaction
32.5ul nuclease free water
20ul BP1 DNA
1ul DNA CS
3.5ul Ultra II end prep reaction buffer
3ul Ultra II end prep enzyme mix
60ul total
20oC for 5min then 65oC for 5 min (in PCR machine - end repair program)

then AMPure Beads resuspend
transfer reaction to 1.5ml eppendorf
add 60ul beads
5 min RT with mix (slow on vortex2 genie)
80% fresh ethanol
pellet  on the magnet
take off sup and wash with 200ul etoh without disturbing pellet twice
then let dry about 30s.
resuspend pellet in 61ul water 2min RT
again pellet on magnits then to new tube for Ligation adapter addition and clean up

60ul cleaned elution of er1 sample
25ul ligation buffer LNB
10ul NEB NExt Quick T4 DNA Ligase
5ul of ligation adapter LA

10min RT

add
40ul AMPure XP 5min mix
5min on slow mix vortex (maybe too fast, hope)
similar cleaning on magnet x2, but wash with SFB (short fragment buffer)250ul  x2,
Weird!  has 2 bands now, after ligations, then made one thick band?  beads effect with molecules??
then dry as before and resusp in 15ul EB elution buffer
get
will want 12ul for Library at the correct concentration, but no Qbit

prepare FCF with
1.17ml
25ul 10mg/ml BSA
30ul Flow Cell Tether
'FCF+' tube

with library in frig.
then check flow cell...
needs to get to 37oC
pores ok?  slow process - just 1 or two? or??  only 6 pores available!??
also computer problems, too many instances of MinION program open?
got a second complete flow cell check and it says still only 6 pores.
Go through the

Then, slide the flow cell priming port cover clockwise to open the port,
then, check for air bubbles and use empty P1000 tip to check continuous liquid
then load 800ul of the priming mix into the flow cell, don't introduce bubbles!
ok
wait 5min before next step.
Mix the LIB (library beads) thoroughly by pipetting.

In a new tube put
37.5ul sequencing buffer (SB)
25.5 ul LIB (just mixed)
12ul of the prepared library (all in 4oC)

Open the SpotON sample port cover (flip open)
load 200ul of priming mix into the priming port

mix the prepared library by pipetting just prior to loading
add 75ul to the flow cell in the sample port ensuring each drop flows into the port.
replace this SPOT on lid and close priming port

at one point saw 9 and even pores ok, but then Yngvar set for 30min run.
we will see.  wash seems hopeless too...
The flow cell was received in September, so just no good now...

No sequence in the half hour. what a waste of a flow cell!
However there are teeny micro bubbles over the seq thing (spot-on, sensor array)...

-To note, the light shield didn't get used...

Should we go through and wash etc do it as an exercise
empty from waste port, then mix up wash
1h at RT,
then empty waste and 500ul of storage buffer across array... and empty waste again.
try to look at pores again, just because?

Yes!
what does it say? warming to 37 again.
just 2 pores in thend... not unexpected...
'below warrantee level'
once more, after re-insert and with light cover on top...

Refrigerator coldness problems always! could be extra reason, for failed nanopores.