Start hood, and clean,

run 15mins and clean again.
Get ice ready...
Make aliquot of clean water in 50ml conical, under hood.
Prepare tubes for the aliquoting, have also plastic bags for aliquots.

  1. When synthesized primers are received, the tubes should be spun in the microcentrifuge to make sure that the lyophilised contents aren't lost upon opening.
  2. The appropriate volume of Sterile water for making a 100uM solution (as written on the shipping documents) should be added. 
  • double check all labels and check off list!!

          (can also use 10mM Tris pH8 for resuspending primers, and these may be more stable - but they might not be suitable for the colorimetric or other methods)

prepare bags
NM- Mammoth N set
NZ- Zhang N set
rA- rAction set
RN- RNaseP set
F3/B3, FIP/BIP, LF/LB
and tubes (label)
and a box for original tubes.

  1. Wait 5 min at RT

  2. Vortex for about 1 min full speed

  3. Wait again 5 min

  4. Check for complete resuspension, can tap and shake

quick spin, to make sure liquid is not in caps, then all to ice.
Aliquot suitable volumes for the application.
Normally, would make about 5+aliquots of the 100uM (as possible)
However, today, for first set of LAMP primers to develop #CoronaDetective
as part of the OpenCovid19 Initiative with JOGL, make just x2
aliquots of all (80-250ul, depending on tube - note on original paper),
plus a  standard mix as below, step 8, into sachets and also
save the initial tubes in a separate freezer box.
Store at -20oC - big freezer special drawer. (took pic)

7) to make inventory page. doc
To note: there are a few primers that had quite low yields, only about 15-18nmols!
(rA-LB, NM-B3 and NZ-LF)
and another with only 33.7nmols (RN-F3)

Should probably write Eurofins and ask for more already??

(at least use smallest amounts of F3/B3 in these reactions...)
however, the official 'synthesis scale' is 0.01umol, so can't complain?
(except FIP/BIP was 0.05umol synths)
See what Guy and Fran think...

  1. optional * done today.

Can prepare a mix for 35 reactions of 25ul of the usual LAMP primer mix, including all  primers (FIP/BIP/F3/B3/LB/LF), is made as follows, per 25ul reaction:

0.4ul FIP and BIP (35 reactions: 20 µl)

0.2ul LB and LF (not always there; 10 µl for 35 reactions)

0.1ul F3 and B3  (5 µl for 35 reactions)
Final Volume: 70ul
So,
this comes out to
28uM FIP/BIP
14.2uM LB/LF
7uM F3/B3

1.4ul per 25ul reaction gives final concentrations of about
1.6uM FIP and BIP,  0.8uM LB and LF, and 0.4uM F3 and B3
(in some cases this can probably be adjusted *assay optimisation)

9)  place the quality certificate datasheets in a safe place (clear lab folder in desk)