Notes de Rachel/Covid-19 LAMP tests (46 notes)/01052020 Trying old school Bst with its own buffer on RNaseP set
Notes de Rachel / Covid-19 LAMP tests (46 notes)
To find out if the old school Bst will show activity with its own buffer.
Master Mix Tubes (for 0,0,R+,R+)
1A - Bst alone
1B - Bst+RTx
1C - Bst+RTx + extra loop primer
2A - Bst2.0 alone
2B - Bst2.0+RTx
2C - Bst2.0 +RTx + extra loop primer
Need to make sure whether I have still 30ul of RNA for the last ones! & decide
(if I have 2 strips of 12 tubes each, but use round top lids, separate closure for neg controls first, before bringing out RNAs!)
As yesterday - clean well with bleach and etoh before anything.
and liberal use of gloves!
For Master Mixes (1A to 2C - 6 tubes)
for each 25ul reaction (x4.5 for all, so no worries about running low...)
2.5ul 10x buffer ThermoPol for oldschool Bst Isothermal for Bst2.0 (11.25 for 1A-2C - 3+3 tubes)
1.4ul* standard or +loop primer mix (6.3 for 1A/1B and 2A/2B - 4 tubes standard - extra loop for Cs 2tubes
extra loop primer - make up with 0.8ul LB and LF in master mix for 4.5 rxns))
1ul extra MgSO4, 100mM (for 4mM more - 6mM final in reaction) (4.5 for all tubes)
3.5ul dNTP, 10mM (15.75 for all 6 tubes)
1ul Bst2.0 (8units) or BstOldSchool (4.5)
1ul RTx (15units) (4.5 for B and C tubes 4)
2.0 ul SYBR-safe diluted 1/250 from stock solution, so final concentration is slightly decreased (0.00032), vs last time (1/2500=0.0004). (9 for all master mix tubes)
water + sample to 25ul final (say 2.5ul or RNA or water per reaction)
25-2.5-(2.5+1.4+1+3.5+2) and (As 1) (B/Cs 2)
water for As - 13.9 (x4.5) =62.6
water for B - 12.9 (x4.5) =58.1
water for Cs - 56.5
do 64 degree incubation again today - but Curtis et al did theirs at 60 degrees!)
to take pictures manually with CCD of ABI Prism7000 machine, as still no real run possible
(every 3-5minutes using stop watch pausing etc...)
can also use UV to look at reactions (transilluminator)
Results:
the old school Bst was quite a bit slower, but stayed clean in 0 tubes to the end, and also seems like there is no need for RTx...
The extra loop primers did not help either.
but the Bst2.0 was a mess... first one to come up was at about 30min and with water input... !!
Most likely the RNA input was degraded after x3 freeze/thaws, so that many fewer copies were there to be amplified - but this still seems like a good primer set!
To get data in Fiji for graphing again, but end point pic with trans-illuminator shows up aok...
(Hope for calibrating the big beast, so normal runs can be done !?)