Notes de Rachel/Covid-19 LAMP tests (46 notes)/28102020 CoronaDetective on blinded samples swabs and saliva

From Hackuarium
Revision as of 20:26, 18 November 2026 by obsidian>Import Obsidian Rachel (Import des notes de Rachel)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Notes de Rachel / Covid-19 LAMP tests (46 notes)

To find out if any of the three unknown samples (the swabs and saliva for each) extracted yesterday will be NM positive in the test.

Make master mix for 16 reactions
1,2 waters, 3,4  extract A1, 5,6 A2 extract, 7,8  extract B1
9,10 extract B2, 11,12, extract C1, 13, 14 extract C2, 15, 16 - BEI RNA 200(14.7dil)
A-C1 -swab extracts
A-C2 -saliva extracts

Hope there is still enough RNAsin!
just dregs - around 1ul? rinsed out tube with water...

For Bst2.0 +RTx 'Corona Detective' reactions make master mix:
for each 20ul reaction.   (master mixes  x17)
2.0ul 10x buffer           (34ul)             
1.12ul* tagged primer mix   NM and RP multiplex primers and quenchers.   (19.4 each)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction)   (17)
2.8ul dNTP, 10mM      (47.6)
0.8ul Bst2.0 (8units)  (13.6)
0.5ul RTx (7.5units)    ( 8.5)
0.5ul RNAsin. (8.5)
water + sample to 20ul final   (do 4ul of extract or control/water per reaction)
16-(all components above with doubleprimers) = 16-9.84=     6.16x17= 104.72
NM fluor primer mix _FAM-NM*FIP + RNaseP fluor primer mix. _hex-RP*LF  
Both quenchers 8.8ul for NM-FIP quench +4.4 for RP-LF quench =13.2ul
20*17=340 total final volume - 68 (samples/water)=272
minus quenchers =258 - (9.84*17=167.28)= 90.7 water

For pcr machine - start with 10min at 55 degrees, to give RTx a better chance to work (since low!)
For NM: 1.6uM for FAM tag FIP NM, so 2.4uM for quencher, 0.8uM for LF/LB, 0.2uM for F3/B3
For RNAseP: LF is the HEX tag, so 0.8uM LF and 1.2uM for the quench... 
base on 'ordinary double loop' mixes...
1.6uM FIP/BIP -     20 µl
0.8uM LB/LF -       10 µl 
0.2uM F3/B3 -         2.5 µl each
                               +5 ul water

Results
started at 2:30, at about 3pm only B1 tubes with first hints of signal?
at 3:25 some tubes came up well - waters and A1s neg, but A2 have RP*hex and one B1 has it too
then in the second row, one B2 is positive, both C1 neg and C2 positives RP*hex - not green!  BEI RNA came up nicely by now, nice and green.  NO problemos!
give another 20min anyway... for endpoint.
(was awesome that none of the 3 looks NM+ at that point! 40min at least
BUT!!!!
next exposure in about 20 more minutes had A1 swabs positive!  (but not saliva)
call it 55min...
then do 10min more...
a saliva in A also came up - results to lab notes
(got letters 'unblinded' and A person will have quarantine and get an official test - already scheduled for Friday). B was the person negative by an official test done just the day before, so this is already a good validation for Corona Detective.  (C is happy to have this result, but it does not really change anything, except they were dizzy and off work on the Monday.)
The swab worked well with this simple kit extraction (better than saliva! for which only one really looks NM positive). We will see if official test is further validation.
In fact: the qpcr result was negative, so a few extra days of quarantine was the worst of this episode - except, as this could be the first late false positive??  but- made the control N plasmid in this lab so... that wasn't a super plan, still... (i.e. just a cross contam, not real false pos...). we will see!