Difference between revisions of "Notes de Rachel/Sam S expts (20 notes)/20180807 test for standard LAMP reaction in the heated system"

From Hackuarium
Jump to navigation Jump to search
obsidian>Import Obsidian Rachel
(Import des notes de Rachel)
 
m (1 revision imported: Import des notes de Rachel — lot 6/6)
(No difference)

Revision as of 20:58, 18 November 2026

Notes de Rachel / Sam S expts (20 notes)

To use the old reaction solutions left for Sam in Freezer earlier in summer

from July 6th
put

DR/serp3
12x 50ul positive reaction mixes
(include extract)
 and 
4x 50ul negative reaction mixes
(only water)

see what is left and try it today
will need to stay by the machine...

Values obtained (written on paper)

Before run, machine empty
B            G         
117          7
empty chip inside
113          10
117          12
107          11
Rinsed chip with water
154          9
146          8
146          9
Then with D+ solution (6July into freezer)
sealed holes with tape (pic)
Before starting run
B               G
151          29
151          28
151          28
Start run (left button of prototype)
B               G       t(secs)
151             27       30
143            17        181
140          14          210
138          13          223
136          11          370
136          12          410
132          12          490 (worry about air bubble?)
131          11 .        535 . (quick peek, not obvious)
128          11          890
127           11           1012
134          13          1056
134          13          1100
134          14          1130
134          16          1160
134          16          1175
134          17           1200
135          18          1250
134         18            1290
134          18        1336
135          18          1380
135          18          1415
135          19          1445
135          19          1500 . (looked again, teeny bubble?)
134          18          1581
133          18          1630
136          18          1685 (look again)
136           18           1705
136          18           1752
135         18 .      1792
134          10         1808 fan started horrid big noise         
 141          16         stop for last reading...

made chart in excel
not impressive!
https://drive.google.com/open?id=19189k4G3FE4oQiG48HLIRb4Ddi4FP9Dd

Got pic of chip to show max bubble
https://drive.google.com/open?id=1xFH7XTKUSEf-oIgEzcIh_OLZQl_b8vvG

(but think was at hole side, under tape... during run)

kept chip at RT then did another round to see if anything else came up 
153  24   start
141   17    1657
143  17     1800 (off for big noise!)

Saved solution and put in tube to test T16 fluor level
also do 2 neg and 2 pos reactions...

Look like normal S.epi T16 run.  solutions still good from freezer...
but chip reaction is only up to about 2000rfu
heating was sub-optimal... most likely. (loose chip -seating not firm or material problem - see pics, wet chip: https://drive.google.com/open?id=1wjrLznzwjSgS2W41UF3MXI9y5TG2iGY2
dry chip: https://drive.google.com/open?id=1HBqw-glQcLOS6EGkGNHwN5ISX_ku-pGq
)

After full run was complete, could see that the maximal signal did decrease with time.
https://drive.google.com/open?id=1915SJPwGEMJ7-b7H7ZlNkqYIzt2AskbF

but they do not decrease to the same level as that of the solution taken from the reacted chip, as seen readily with the overlay.
https://drive.google.com/open?id=1025CnmJffXSMWbVhnZa8We2HJan4Up5n

Perhaps the low rfu is due (also) to bleaching...

the machine can maybe be set to do shorter excitation?
maybe the detection wavelength is not optimal for the evagreen??

This time in DR conditions, Green signal went from only 9-19
vs last time 9-90 (with FAM and quench reaction)