Difference between revisions of "Notes de Rachel/Covid-19 LAMP tests (46 notes)/20211126 Corona Detective mastermix -stability at 4oC"
obsidian>Import Obsidian Rachel (Import des notes de Rachel) |
m (1 revision imported: Import des notes de Rachel — lot 1/6) |
(No difference)
| |
Revision as of 20:54, 18 November 2026
Notes de Rachel / Covid-19 LAMP tests (46 notes)
To test if a mastermix with all inside will be stable in refrigerator.
make up 3x master mixes
A) usual NEB Master mix
2.0ul 10x buffer - Isothermal buffer for Bst2.0 (x27 for all components for the 9x3 reactions) (54ul)
2ul 10x primer mix (54ul)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) 27ul
2.8ul dNTP, 10mM (75.6ul)
0.8ul Bst2.0 (8 units) (21.6)
0.5ul RTx (7.5 units) (13.5 ul) For Guy's master mixes he uses 6units RTx & 6.4units Bst2 per reaction
water + sample to 20ul final (say 5ul of sample per reaction)
WATER for rxns with 7mM and RTx ) =5.9 x27 = 159.3
distribute 15ul per rxn
B) 2xNEB mix - so 15ul per reaction (5ul sample addition)
270ul 2x soln,
54ul 10x primers
96ul water
Do I have enough?!? looks like especially need more 10x primers!
10X primers in 100ul with 100uM stocks
Fam-FIP/BIP 16 Micromolar (µM) - 16 µl each
LB/LF 8 Micromolar (µM) - 8 µl each
F3/B3 2 Micromolar (µM) - 2 µl each
Anti-FIP-Q 24 Micromolar (µM) - 24 µl
24 µl DNAse/RNAse free water
After master mixes made A and B vortex and spin. A is more bubbly, B goopier
change gloves, get fresh tubes and make x3 rows of each 15ul.
(water to first two of today's A and B)
after extra strips sealed and to frig, dilute control virus.
to run reactions today,
water x2
2copies x2
20 copies x2
200 copies x2 (fresh v2000 from 1/100 dil and serial dils )
A row and B row
plus x2 singlet tubes A0 and A20...
(should order more 8-tube strips!)
63oC first pics at 30min
one tube of 200 copies lit up in both strips, but after an hour and an hour and a half, still the same!
continue!
could my primer not be so good now? can start from new aliquots or original tubes for all next time?
plenty of enzymes.
maybe what is the problem??
by the end, at least got both 200s up in A, but still only 1 of them for B.
looks like 20 copies might come? could give another half hour?? -almost 7pm on Friday night though! put to frig and do more incubation monday?!
29nov21
have 2 strips left of unused A and B mixes.
have a feeling that I did 1/100 serial dilutions for the virus! So, the '200' tubes maybe really had 20 copies! today, to do reactions with these 3day old master mixes and fresh viral suspensions with serial 10x dils!
Also, wondering if extra B3 might help for more efficient reverse transcription.
So: A1 and B1 as usual and A2/B2 with extra 0.1uM of B3 primers
made 1/100 dil of B3 (so 1uM) and used 2ul in the 20ul rxn (with 3ul h2o)
add to all 16 '2' tubes...
Then water to all 0 tubes and sealed,
then water to v tubes and dilution of fresh virus aliquot,
then distribute 2,20,200 copies to tubes.
incubate tubes from last Friday for 15min first pics
no diff
then to 30min...
with all...
First pics of last Friday, no diff.
First 30min pics
A1 nada
A2both 200 copies up.
B1 weird 20 copy bottom green!
B2 nada
no diff for last Friday's tubes,
Second set of pics (1h)
third 1.5h
fourth 2h...
maybe A2 is better - but, late false pos!??
weird that B2 gave nada, however! neb 2x doesn't like extra B3 primer?!!
don't get it