Difference between revisions of "Notes de Rachel/P1Lab Fall2020 (5 notes)/07102020 RNA extractions with Monarch Kits + Multiplex Corona"
obsidian>Import Obsidian Rachel (Import des notes de Rachel) |
(Adding Category:Note (batch edit via API)) |
||
| (One intermediate revision by one other user not shown) | |||
| Line 50: | Line 50: | ||
pics to go in drive | pics to go in drive | ||
| + | [[Category:Note]] | ||
Latest revision as of 09:47, 26 November 2026
Notes de Rachel / P1Lab_Fall2020 (5 notes)
Set up for RNA extraction with the Monarch Kit again, with Tatiana.
Make 3 'saliva samples' (A, B, C) 2
2
2 saliva samples B and R (Benni and RA )
1 saliva spiked with BEI RNA (with RNAsin?).
2 cell RNAs (inner cheek cells)
Follow Monarch kit instructions, note details on sheet.
also following their saliva extraction guidelines (no PK or DNAse! but equal vol of DNA/RNA soln) (protocol from NEB)
https://www.neb.com/protocols/2020/08/14/rna-extraction-from-saliva-using-the-monarch-rna-cleanup-kits
(should have tried an RTx- set too?!)
Details on printed page also for picture to add to drive.
For Bst2.0 +RTx 'Corona Detective' reactions make master mix:
(for each 20ul reaction)
2.0ul 10x buffer (34ul)
1.12ul* tagged primer mix multiplex primers and quenchers. (19.4)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) (17)
2.8ul dNTP, 10mM (47.6)
0.8ul Bst2.0 (8units) (13.6)
0.5ul RTx (7.5units) ( 8.5)
0.5ul RNAsin. (8.5)
water + sample to 20ul final (say 2ul of extract or control/water per reaction
with extra 55degree 10min
with 16 reactions, made master mixes x17
18-(all components above -doubleprimers) 7.86x17= 134
NM fluor primer mix _FAM-NM*FIP + RNaseP fluor primer mix. _hex-RP*LF
Both quenchers 8.8ul for NM-FIP quench +4.4 for RP-LF quench, for water -> 120.5ul
1,2 waters, 3,4 B extract, 5,6 R extract, 7,8 BEI RNA 100
9,10 older cheek cell RNA (R2 from 12.7), 11,12, B extract + BEI RNA100, 13, 14 R extract +BEI RNA100, 15, 16 - N 50 copies. (old dilution)
For pcr machine - start with 10min at 55 degrees for *+* tubes, to give RTx a better chance to work (since low!)
For NM: 1.6uM for FAM tag FIP NM, so 2.4uM for quencher, 0.8uM for LF/LB, 0.2uM for F3/B3
For RNAseP: LF is the HEX tag, so 0.8uM LF and 1.2uM for the quench...
base on 'ordinary double loop' mixes...
1.6uM FIP/BIP - 20 µl
0.8uM LB/LF - 10 µl
0.2uM F3/B3 - 2.5 µl each
+5 ul water
great results, but not great timing (no probs for endpoint detection) with JOGL call...
Tatiana needs an extra bit of pipetting attention for small volumes (a couple false negative B tubes)
pics to go in drive