aim:
The goal is to test the chip, with a control (sterile water) and physiological serum (0.9% saline in sterile water).
We will use the chip to test if it can be used for the comet assay. Therefore, 3 experiments will be made, manipulations in sterile water or in physiological serum. For the first case, we will also test the electrophoresis within the chip.
Violaine has harvested her cells in 5 ml of sterile saline and in 5 ml of physiological serum.
Centrifuged for 3 minutes and 4000 rpm. Discard the supernatent and resuspend the cells in 1 ml.
(Note: there is only a few single cells. 30 cells / square for both sterile and physiological)
LMP agarose 1.5% made with sterile water and with physiological serum (0.15g in 10 ml).
100 ul of LMP agarose for each pad.
Gently pour the pad with a paintbrush into the chip, on the glass lid and then put 600 ul of 0.9% saline in each well.
--> 9 wells, we need 5.4 ml of saline 0.9%
For the H2O2 treatment, pour 600 ul of H2O2 0.5% and let it for 5 minutes.
--> For 1 well, we need of 100 ul of H2O2 3% in 500 ul of 0.9% Saline (respectively sterile or physiological). We have 3 wells, we need ~1,8 ml of saline
L physio
R sterile
12 wp sterile
Wash 5 times with saline (sterile or physiological respectively).
--> 3 wells, we need
For the PK treatment:
Pour 600ul of PK Buffer, twice and let for 5 minutes.
--> 3 wells filled twice per chip, we need therefore 3.6 ml of PK Buffer.
--> Finally, 7.2 ml of Sterile PK Buffer and 3.6 ml of Physio PK Buffer
I will make 8 ml and 8 ml.
For the PK+ treatment:
Pour 600ul of PK+ and let overnight.
--> 2 wells per chip (treated and normal), we need therefore 1.2 ml of PK+.
--> Finally, 2.4 ml of Sterile PK Buffer and 1.2 ml of Physio PK Buffer
_______________________________________________________________________________________________
PROBLEMS:
- The wells leaked and everything went out. We don't know why for certain, but maybe:
- The volumes were too big. We put the exact volume of the wells, in order to fill them completely; maybe the liquids inside them communicated and leaked through the holes of the electrophoresis.
- Solution: Put less volume
- The material is porous and the liquid can leak through it.
- Solution: Find another material
- PK Buffer dissolves the glue of the tape used to seal the holes of the wells.
- Solution: Find another way to seal the holes of the wells OR make no holes in the wells and find another way to remove the glass lids or the pads.
- Maybe not this hypothesis, the scotch seems to have been well sealed and did not let liquid leak.
- The soapy water has a higher surface tension than the water and thus goes into all wells
- The volumes were too big. We put the exact volume of the wells, in order to fill them completely; maybe the liquids inside them communicated and leaked through the holes of the electrophoresis.
- Test of the chip, with only water and soapy water
- Quickly, soapy water leaks out of the chip (whereas simple water stays).
- So certainly, it is because soapy water has a surface
- Test of the chip, with only water and soapy water
- Quickly, soapy water leaks out of the chip (whereas simple water stays).
- So certainly, it is because soapy water has a lesser surface tension than simple water and therefore it can "jump" above the walls of the wells.
- Test of the electrophoresis on the chip:
- Sealing
- Scotch
- Put scotch on all the edges (between the glass plate and the chip) and then the liquid (here water, to test the seal).
- Not enough sealing, the water can leak in the space in between the two plates.
- Parafilm
- Put parafilm, no glass slide
- Sealing complete but need some flat surface underneath to flatten the parafilm
- Put the chip on another bloc, where we can put the liquid and even if there is leak, the chip is constantly surrounded by electrophoresis liquid:
- (Photo of the setup)
- Try to put the current on
- Note: Find a way to keep the wires immobile in the chip when moving them AND to not let bubbles inside the chip
- Results: It works quite well! There is bubbles on one of the wire