Notes de Rachel / Covid-19 LAMP tests (46 notes)
to see if we are good to go! :)
NM + RNAseP multiplex even...
in first trials today will do*
1.6uM for FAM tag FIP NM, so 2.4uM for quencher, 0.8uM for LF/LB, 0.2uM for F3/B3
For RNAseP, LF is the HEX tag, so 0.8uM LF and 1.2uM for the quench...
base on 'ordinary double' loop mixes...
1.6uM FIP/BIP - 20 µl
0.8uM LB/LF - 10 µl
0.2uM F3/B3 - 2.5 µl each
+5 ul water
maybe should do repeat of trial 8 for the machine???
protK before saliva addition? No, Simpler (don't use up all enzyme now!)
2 master mixes x8 (1,2 - water, 3,4 BEI RNA 20, 5,6 BEIRNA 200, 7,8 sample x (swab, protK saliva? cell RNA with Tatiana?? - in the end Saliva for both 7,8 and N5 DNA control for sample 8 also)
A BO +RTx with NM (each enzyme with its correct buffer)
B Bst2.0+RTx with NM
Master Mixes for realtime amp data:
2.0ul 10x buffer - Isothermal buffer for Bst2.0 (x9 for all components for the 8 reactions) (18ul)
1.12ul* primer mix (10.1ul)
1ul extra MgSO4, 100mM (for 5mM more - 7mM final in reaction) (9ul)
2.8ul dNTP, 10mM (25.2ul)
0.8ul Bst2.0 (8units) or BstLarge Fragment (7.2)
for both master mixes RTx (7.2)
1.6 ul SYBR-safe diluted 1/250 from stock solution. (14.4)
water + sample to 20ul final (say 2ul of RNA or water per reaction)
WATER = 7.9 x9 = 71.1
Master Mixes for first trials with quenchers:
To do A and B mixes for all 3 of these, to just incubate and look on fluor detector?
(to check how they look on the machine too... cold vs hot should be big differences)
Will add Quencher to the master mix, not the primer sets... No SYBR SAFE in the 1,2,3 master mixes, so add 14.4ul more water
1) NM fluor primer mix _fam-FIP for 2.4uM of quencher / rxn. 4.4ul 100uM quench stock will be added to the 180ul final vol master mix and water at 66.7->81.1
2) RNaseP fluor primer mix. _hex-LF for 1.2uM quench/rxn 2.2 ul in mastermix and water at 68.9 -> 83.3
3) both of them together! fam and hex then water decreases to 54.4ul -> 68.8
then use as above (will need to adjust water)
So, 8 master mixes in the end, all BO vs Bst2 comparisons: 2 for prism run (and put the residual master mix for all also in a strip row), and 6 just incubated at 65 degrees, with pictures on the 'GMO detective' detector every 10 minutes
For sample tubes 7 and 8, especially want to look further into the saliva issue.
Got Tatiana to collect around 250ul of saliva: 95degrees 5min, then 2ul ProtK (mol bio grade) at 37 degrees for 10min (a big more), then again 95degrees 5min and pelleted 2min - collected supernatent. and put 2ul in both tubes 7 and 8. To tube 8 also added 1ul of N plasmid ('5' copies per microliter)
Again- Approximately Isothermal conditions, so get good traces:
65 2min, 64 1min, then 'cycle' 10s 65 and 1:50s 64. (for imaging every 2 min)
do 35 cycles again with dissociation protocol clicked and 60 degrees...
Results:
The FAM/quench works really well for the NM primer set! B2 came up faster than BO (by 20 min vs by 30min), but both seem really good.
It has to cool down as expected to see the negatives vs positives.
The HEX did not give anything on the DIY detector, and there wasn't proper mRNA for the amplification test, and maybe it wouldn't even be expected to show up. (on the prism might be more promising, or UV light and filter/black light? might work...) What was worrisome was that with the Bst2 conditions, there was huge fluorescence in all the tubes with the combination of Hex and Fam tags, that was not observed in BO buffer. Maybe there is a competition for one or the other quencher in the mix?
-> adding extra quench to tubes 2/4/6/8 did reduce this apparent fluor, but didn't save the reactions...
(put 1ul quenchNM-FIP and 0.5ul quenchRP-LF and 5ul water, then 1ul into the even tubes with B2 buffer- made a difference in fluor at least!)
The treated saliva was not promising or helpful... didn't get amplified with the RNAseP set and and no amplification of the N5 occurred in tube 8s even - blocking rxn? but maybe the 5copies/ul dilution of the N plasmid was old or just not enough? or maybe the second 95deg didn't kill off all the resid PK?
On the prism, the 20/200 RNAs amplified very well, not sure if saliva killed the N5 from coming up, or if N5 was just not good anymore...
As usual BO reactions with water stayed clean to end, but Bst2 ones came up late.
bit worried that the endpoint tifs aren't exactly in focus now - maybe camera got knocked when lab rearrangments were done?
Also
Saved all tubes at RT to look again. seemed very stable overnight.
Need proper cell mRNA for the test next too.
SHOULD re-fix protocol as done - with water adjustments, Quench added to Master mixes, etc!