Difference between revisions of "Notes de Rachel/Sam S expts (20 notes)/20180807 test for standard LAMP reaction in the heated system"
m (1 revision imported: Import des notes de Rachel — lot 6/6) |
(Adding Category:Note (batch edit via API)) |
||
| Line 107: | Line 107: | ||
vs last time 9-90 (with FAM and quench reaction) | vs last time 9-90 (with FAM and quench reaction) | ||
| + | [[Category:Note]] | ||
Latest revision as of 09:52, 26 November 2026
Notes de Rachel / Sam S expts (20 notes)
To use the old reaction solutions left for Sam in Freezer earlier in summer
from July 6th
put
DR/serp3
12x 50ul positive reaction mixes
(include extract)
and
4x 50ul negative reaction mixes
(only water)
see what is left and try it today
will need to stay by the machine...
Values obtained (written on paper)
Before run, machine empty
B G
117 7
empty chip inside
113 10
117 12
107 11
Rinsed chip with water
154 9
146 8
146 9
Then with D+ solution (6July into freezer)
sealed holes with tape (pic)
Before starting run
B G
151 29
151 28
151 28
Start run (left button of prototype)
B G t(secs)
151 27 30
143 17 181
140 14 210
138 13 223
136 11 370
136 12 410
132 12 490 (worry about air bubble?)
131 11 . 535 . (quick peek, not obvious)
128 11 890
127 11 1012
134 13 1056
134 13 1100
134 14 1130
134 16 1160
134 16 1175
134 17 1200
135 18 1250
134 18 1290
134 18 1336
135 18 1380
135 18 1415
135 19 1445
135 19 1500 . (looked again, teeny bubble?)
134 18 1581
133 18 1630
136 18 1685 (look again)
136 18 1705
136 18 1752
135 18 . 1792
134 10 1808 fan started horrid big noise
141 16 stop for last reading...
made chart in excel
not impressive!
https://drive.google.com/open?id=19189k4G3FE4oQiG48HLIRb4Ddi4FP9Dd
Got pic of chip to show max bubble
https://drive.google.com/open?id=1xFH7XTKUSEf-oIgEzcIh_OLZQl_b8vvG
(but think was at hole side, under tape... during run)
kept chip at RT then did another round to see if anything else came up
153 24 start
141 17 1657
143 17 1800 (off for big noise!)
Saved solution and put in tube to test T16 fluor level
also do 2 neg and 2 pos reactions...
Look like normal S.epi T16 run. solutions still good from freezer...
but chip reaction is only up to about 2000rfu
heating was sub-optimal... most likely. (loose chip -seating not firm or material problem - see pics, wet chip: https://drive.google.com/open?id=1wjrLznzwjSgS2W41UF3MXI9y5TG2iGY2
dry chip: https://drive.google.com/open?id=1HBqw-glQcLOS6EGkGNHwN5ISX_ku-pGq
)
After full run was complete, could see that the maximal signal did decrease with time.
https://drive.google.com/open?id=1915SJPwGEMJ7-b7H7ZlNkqYIzt2AskbF
but they do not decrease to the same level as that of the solution taken from the reacted chip, as seen readily with the overlay.
https://drive.google.com/open?id=1025CnmJffXSMWbVhnZa8We2HJan4Up5n
Perhaps the low rfu is due (also) to bleaching...
the machine can maybe be set to do shorter excitation?
maybe the detection wavelength is not optimal for the evagreen??
This time in DR conditions, Green signal went from only 9-19
vs last time 9-90 (with FAM and quench reaction)