Difference between revisions of "Notes de Rachel/Interns, Summer 2018 (51 notes)/20180821 Comet assay sterile vs bidistilled water and SYBR Safe vs Methylene blue"

From Hackuarium
Jump to navigation Jump to search
obsidian>Import Obsidian Rachel
(Import des notes de Rachel)
 
(Adding Category:Note (batch edit via API))
 
(One intermediate revision by one other user not shown)
Line 118: Line 118:
 
Maybe the comets were/seemed longer in the bidistilled thing because the water still contained ions and helped the DNA to migrate further
 
Maybe the comets were/seemed longer in the bidistilled thing because the water still contained ions and helped the DNA to migrate further
  
 +
[[Category:Note]]

Latest revision as of 09:39, 26 November 2026

Notes de Rachel / Interns, Summer 2018 (51 notes)

New materials produced:

  • Saline 0.9% with sterile water (0.45 g in 50 ml of sterile water)
  • Agarose 1.5% (0.375 g of agarose in the previous saline solution, with sterile water)
  • Notes:
    • We already had agarose 1.5% from a previous experiment
    • Sterile water: made by reverse osmosis

Cell counting:

  • sterile water: (26 + 37) equals 31 cells per square in average --> 310'000 cells in the sterile water falcon
    • Need of 20'000 cells per tube --> 15 tubes for the sterile water
    • Therefore, 1 ml/15 equals 62.5 ul per tube
  • Bidistilled water: 25 cells per square --> 250'000 cells in the bidistilled water falcon
    • Need of 20'000 cells per tube --> 12 tubes for the bidistilled water
    • Therefore, 1 ml/12 equals 83.3 ul per tube

Preparation of the pads:

  • Configuration of the pads: https://docs.google.com/spreadsheets/d/1mxpHjNSaEs1Flz8p0s5VI9-mn_Sjh86txncNkoV5YVE/edit#gid=1625783844
    • Water first: 
      • In the two first columns, 500 ul of sterile water were put
      • In the two last columns, 500 ul of bidistilled water were put
    • Treated line: 
      • 17 ul of H2O2 3% put in the wells of the treated line, that had already 500 ul of saline --> 0.1% of H2O2
      • This liquid was discarded (no bubbles seen)
      • Washed twice (1 minute with the first wash) with saline water (sterile and bidistilled for the respective wells)
    • Normal and treated lines
      • Their 500 ul of saline have been discarded and 400 ul of PK buffer (sterile and bidistilled) have been put twice in the wells (sterile and bidistilled respectively)
      • 75 ul of Proteinase K put in 1.6 ml of Proteinase Buffer Sterile
      • 75 ul of Proteinase K put in 1.6 ml of Proteinase Buffer Bidistilled
      • 400 ul of each of these solutions were put in the wells at 13:13 at room temperature, until 14:38
      • The media were discarded and 750 ul of alkaline TBE was put for 5 minutes in the 8 wells (4 normal and 4 treated). This was done 2 times (2 washes)
    • For all wells:
      • 4 ml of 5X TBE were put in 16 ml of bidistilled water
      • All media discarded
      • 750 ul of this solution was put in each well for 5 minutes. This was done 2 times (2 washes)
    • Put TBE 1X as the medium for the electrophoresis
    • Put 4 pads on glass plates (from left to right): 
      • 1st plate: 4 pads Sterile Normal (SYBR Safe and Methylene blue differenciation will be made after)
      • 2nd plate: 4 pads Bidistilled Normal (SYBR Safe and Methylene blue differenciation will be made after) Photo on whatsapp
      • 3rd plate: 5 pads Sterile Treated (SYBR Safe and Methylene blue differenciation will be made after) Photo on whatsapp
      • 4th plate: 5 pads Bidistilled Treated (SYBR Safe and Methylene blue differenciation will be made after) Photo on whatsapp
    • Electrophoresis: 12V for 18 min, 5 mA
    • Legends:
      • USS: Untreated Sterile SYBR Safe;
      • NSS: Normal Sterile SYBR Safe;
      • TSS: Treated Sterile SYBR Safe;
      • UBS: Untreated Bidistilled SYBR Safe;
      • NBS: Normal Bidistilled SYBR Safe;
      • TBS: Treated Bidistilled SYBR Safe
    • Box 1: 3 pads TSS and 2 pads NSS;
    • Box 2: 1 pad USS, 1 pad UBS, 3 pads TBS and 2 pads NBS
    • Methylene blue staining:
      • Put Methylene Blue
      • Rince
      • Put Iodine
      • Wash with 1 ml of bidistilled water
      • Wash with 1 ml of TBE



Observations:

AT UNIL

  • For TSS
    • Many comets, with small tails
      • Note: next time, put more H2O2 in order to cover the pad.
    • Put 20'000 cells per pad instead of 2’000 but the result is good too
      • But a lots of clumps, maybe 10’000 would be better
    • While using the application on the computer next to the microscope, mind clicking the image before saving to ensure that the image saved is the one you wanted.
    • In the middle pad there are lots of cells so we could try to put methylene blue on it
  • For TBS plus de scotch
    • We observe comets
      • Work with bi-distilled water (from the Micropad startup) => No need of sterile water 😄
      • Maybe test the distilled water from Migros (eau pour le fer à repasser)
  • For NBS (2 pads)
    • We observe some tails, they seems even longer than the ones from TBS
  • For NSS (2 pads) scotch se barre
    • We observe tails that seems longer than the previous ones
    • NSS looks better the NBS => maybe sth in the bi-distilled water breaks the cells. The comets seems smaller in NSS than NBS (visually)

Need to compare UBS and NSS to ensure that the bi-distilled water has an impact on the cell

  • For UBS (1 pad)
    • Cells look rounds, healthy
    • GFP + phase
    • Some looks weird, maybe apoptosis 
  • For USS (1 pad with bubbles)
    • Cells look good

AT HACK

After a night in the fridge they look still super black/brown

  • But slowly the dark color faint when shaked

Tried to find a filter at ~665nm for methylene blue

  • Results: 
    • The untreated wells give nice cells, where their nuclei is colored
      • But after several hours, the cells seem to shrink.
    • On the other hand, the normal and treated wells do not give anything, we cannot see comets...





NOTES:
Find a method to connect the two slides => maybe with the imprimante
Minimum 5 pictures are needed to make a comparison !
Maybe the comets were/seemed longer in the bidistilled thing because the water still contained ions and helped the DNA to migrate further